Software Release Bulletin


Chemstation     ChemStation (G2070BA)

May 23 2012 5:10AM

Preface
-------------------------
This Software Release Bulletin documents all fixes and enhancements that are incorporated the new release.
-------------------------


Keyword Glossary


35900 Control
7890A Control
ACQ
Data Analysis
Data Analysis , SEQ
Documentation
ECM Integration/OpenLab Option
GC Analysis
GC Autosampler
Initialization
Installation
Integration
Method
Reports
SEQ
User Interface

Keyword Index


Keyword: 35900 Control

35900: Unable to set up Timed Events
KPR#:1707
ADC can be non-responsive after long reprocessing
KPR#:1739

Keyword: 7890A Control

Logbook lists a disconnection from a 7890A as module 'Unknown Ev'
KPR#:1475

Keyword: ACQ

Not possible to modify data analysis parameters of method during a single run in the offline; cannot save
KPR#:1641
Running dual tower sequence with all methods saved as Dual, get “Injector doesn’t match the sequence’s DUAL Injector”
KPR#:1732
Changes in sequence methods made during acquisition are not saved automatically
KPR#:1742

Keyword: Data Analysis

Red X mark displays in data explorer of offline data analysis
KPR#:1706
PDF reports might not be updated when the file is viewed with PDF reader
KPR#:1662

Keyword: Data Analysis , SEQ

PDF reports might not be updated when the file is viewed with PDF reader
KPR#:1662

Keyword: Documentation

Understanding Your ChemStation manual needs more information about peak width calculations
KPR#:1734

Keyword: ECM Integration/OpenLab Option

Problems after moving items from ChemStation to a different LCDF path in ECM
KPR#:1215
chromatogram data not included in the ACAML file when a sequence is acquired in "Acquisition only" mode
KPR#:1453
ChemStore legacy data doesn't get reprocessed automatically
KPR#:1714

Keyword: GC Analysis

Self-assembled result sets do not have full functionality for GC data
KPR#:1741

Keyword: GC Autosampler

Unclear error message: Undefined symbol: License$
KPR#:1738

Keyword: Initialization

Sporadically launch of an online session will not proceed
KPR#:1618

Keyword: Installation

missing link and no automatic installation of the localized readme.txt of the CDS ed.
KPR#:1589

Keyword: Integration

Integration unable to handle complex chromatogram for area sum integration event
KPR#:981
Data lost when using area sum integration event, must have 0.06 minute space between groups
KPR#:1642

Keyword: Method

Method marked Dirty on initial "Specify Report ..." with DEF_GC.M
KPR#:1694

Keyword: Reports

ChemStation does not report resolution as defined in European Pharmacopoeia
KPR#:1011
acquisition parameters missing in the Method section of the sequence summary report when Unique Folder Creation is switched off
KPR#:1440
sample name not reported with Intelligent Reporting in non-result set data that have been acquired prior to C.01.01
KPR#:1679
ChemMain.exe can crash when a long file path is used
KPR#:1703

Keyword: SEQ

PDF reports might not be updated when the file is viewed with PDF reader
KPR#:1662
Mismatch for sequence completed time stamp in sequence and instrument logbook
KPR#:1317
Headspace or PAL Sampler sequence with methods saved to a location other than C:\ does not run
KPR#:1636
C.01.03: Partial Sequence Acquisition Does Not Work with Headspace with the Back Sequence Table
KPR#:1709
Opening the sequence table, click on a cell, make no changes, and select OK causes sequence to stop with 7697A Headspace Sampler
KPR#:1713
Partial sequence with methods saved to a location other than C:\ does not run with headspace and PAL Samplers
KPR#:1736

Keyword: User Interface

Change Access Level... is not available from menu bar
KPR#:1695
The logged-on user is not displayed in lock dialog
KPR#:1740


KPR#: 1707    Product: Chemstation     ChemStation (G2070BA)    C.01.04

Keyword: 35900 Control

One-line Description:

35900: Unable to set up Timed Events

Problem:
In the 35900 configuration, the Timed Events table is populated with default values. 
In the method, with the injection source set to 35900 Pre-Run Events, the Setup ADC Timed Events window opened but would not permit setup of the events. All buttons and drop down menus were gray and inaccessible.
Temporary Solution:
n/a
Fix Information:
 Fixed in revision C.01.04

KPR#: 1739    Product: Chemstation     ChemStation (G2070BA)    C.01.04

Keyword: 35900 Control

One-line Description:

ADC can be non-responsive after long reprocessing

Problem:
The ADC might become NOT READY when the software is busy for a long time (e.g sequence reprocessing with IR to screen). When the system is no longer busy it cannot communicate with the ADC.
Temporary Solution:
Software and instrument need to be restarted.
Fix Information:
Fixed in revision C.01.03.

KPR#: 1475    Product: Chemstation     ChemStation (G2070BA)    C.01.04

Keyword: 7890A Control

One-line Description:

Logbook lists a disconnection from a 7890A as module 'Unknown Ev'

Problem:
Logbook lists a disconnection from a 7890A as module 'Unknown Ev'
Temporary Solution:
n/a
Fix Information:
Fixed in revision C.01.04

KPR#: 1641    Product: Chemstation     ChemStation (G2070BA)    C.01.04

Keyword: ACQ

One-line Description:

Not possible to modify data analysis parameters of method during a single run in the offline; cannot save

Problem:
It is not possible to change integration parameters in the offline for a single run and save the new parameters in the method file.
Temporary Solution:
Data analysis parameters can be updated and saved if using a sequence.  It is recommended to prepare a single run sequence.
Fix Information:
 Fixed in revision C.01.04

KPR#: 1732    Product: Chemstation     ChemStation (G2070BA)    C.01.04

Keyword: ACQ

One-line Description:

Running dual tower sequence with all methods saved as Dual, get “Injector doesn’t match the sequence’s DUAL Injector”

Problem:
1. Create seq table, with one run (Method A) for the front table, one run (Method B) for the back table this must be done first
2. Before line one of the sequence completes, add a run2 to the back table only (BACK.M) by append
3. When the back run is in process, add a sequence line to the front table, use a different method (method A) 
4. Message stating that Injector doesn’t match the sequence’s  DUAL Injector occurs event though all methods are saved with Dual injection selected as the injection source
Temporary Solution:
n/a
Fix Information:
Fixed in Revision C.01.04.

KPR#: 1742    Product: Chemstation     ChemStation (G2070BA)    C.01.04

Keyword: ACQ

One-line Description:

Changes in sequence methods made during acquisition are not saved automatically

Problem:
Run a sequence and change the method during the acquisition (e.g. change the runtime). There is no possibility to save the method. The method should get saved at the end of the run automatically, but it does not. This issue occurs in C.01.01 - C.01.03 versions.
Temporary Solution:
n/a
Fix Information:
Resolved in Revision C.01.04.

KPR#: 1706    Product: Chemstation     ChemStation (G2070BA)    C.01.04

Keyword: Data Analysis

One-line Description:

Red X mark displays in data explorer of offline data analysis

Problem:
A red X mark displays in data explorer of offline data analysis during sequence.
Temporary Solution:
During data acquisition, don't display the subdirectory in the offline data explorer window.
Fix Information:
 Fixed in revision C.01.04

KPR#: 1662    Product: Chemstation     ChemStation (G2070BA)    C.01.04

Keyword: Data Analysis , SEQ

One-line Description:

PDF reports might not be updated when the file is viewed with PDF reader

Problem:
A pdf report file file cannot be updated when it is open in Acrobat reader or any other PDF viewer. It can also not be uploaded to ECM while it is open.
Temporary Solution:
Close PDF viewers before reporocessing or printing reports.
Fix Information:
C.01.03 comes with a report file viewer that works on a copy of the report file.

KPR#: 1734    Product: Chemstation     ChemStation (G2070BA)    C.01.04

Keyword: Documentation

One-line Description:

Understanding Your ChemStation manual needs more information about peak width calculations

Problem:
The manual needs updated to include more information on the peak width calculation. Is the peak width calculation meant to approach, approximate, or enhance the standard half-height method of calculation.
Temporary Solution:
n/a
Fix Information:
n/a

KPR#: 1215    Product: Chemstation     ChemStation (G2070BA)    C.01.04

Keyword: ECM Integration/OpenLab Option

One-line Description:

Problems after moving items from ChemStation to a different LCDF path in ECM

Problem:
If a ChemStation master method, sequence template, or data set is downloaded from ECM and afterwards the copy in ECM is moved to a different LCDF path, the local copy in the ChemStation will not be linked to the item in ECM anymore. This will lead to errors during automatic data upload, local method/sequence update from ECM, viewing the ECM Properties, etc.
Temporary Solution:
n/a
Fix Information:
Fixed in revision C.01.04

KPR#: 1453    Product: Chemstation     ChemStation (G2070BA)    C.01.04

Keyword: ECM Integration/OpenLab Option

One-line Description:

chromatogram data not included in the ACAML file when a sequence is acquired in "Acquisition only" mode

Problem:
The chromatogram data not included in the ACAML file when a sequence is acquired in "Acquisition only" mode.
Temporary Solution:
n/a
Fix Information:
Fixed in revision C.01.04

KPR#: 1714    Product: Chemstation     ChemStation (G2070BA)    C.01.04

Keyword: ECM Integration/OpenLab Option

One-line Description:

ChemStore legacy data doesn't get reprocessed automatically

Problem:
Migrated ChemStore data acquired with ChemStation rev. A..01.03 and earlier does not get automatically reprocessed when downloading the files from ECM.
The automatic reprocessing is requried to populate the ChemStation navigation table. The reprocessing stops with error message: 
Sequence was aborted due to a file access error. Please restart your sequence.

Restarting the sequence reprocessing does not resolve the error.
Temporary Solution:
Affected sequences cannot be reporcessed automatically, but the runs are available as single runs.
Re-calibration using the calibration samples is possible interactively. Result review can be automated by using the auto-stepping feature of recalculation mode.
Fix Information:
This defect will be fixed in ChemStation rev. B.04.03 Service Pack 1 as well as rev. C.01.04.

KPR#: 1741    Product: Chemstation     ChemStation (G2070BA)    C.01.04

Keyword: GC Analysis

One-line Description:

Self-assembled result sets do not have full functionality for GC data

Problem:
A result set that is built from individual runs does not provide the full functionality for GC runs. The resulting sequence does not support front or back injections. Reprocessing is not possible, because there is a mismatch with the method.
Temporary Solution:
N/A
Fix Information:
Fixed in revision C.01.04.

KPR#: 1738    Product: Chemstation     ChemStation (G2070BA)    C.01.04

Keyword: GC Autosampler

One-line Description:

Unclear error message: Undefined symbol: License$

Problem:
When a GC method is run that uses a vial number not valid for the configuration, an error message "Undefined symbol:  License$" appears.
Temporary Solution:
The error message is misleading. Check the vial range whether it makes sense for the selected injection source.
Fix Information:
Fixed in revision C.01.04.

KPR#: 1618    Product: Chemstation     ChemStation (G2070BA)    C.01.04

Keyword: Initialization

One-line Description:

Sporadically launch of an online session will not proceed

Problem:
Sporadically the launch of an online session will not proceed in a networked setup.
Temporary Solution:
Launch the online session a second time. The second attempt usually works.
Fix Information:
No fix planned

KPR#: 1589    Product: Chemstation     ChemStation (G2070BA)    C.01.04

Keyword: Installation

One-line Description:

missing link and no automatic installation of the localized readme.txt of the CDS ed.

Problem:
The localized readme.txt of the CDS ed. are present on the root of the CD/DVD holding the software (OpenLAB CDS ChemStation ) .
Nevertheless for the localized readme.txt  no automatic installation of the file to the harddisk will take place, the link in the programs folder links to the english readme.txt.
Temporary Solution:
n/a
Fix Information:
 Fixed in revision C.01.04

KPR#: 981    Product: Chemstation     ChemStation (G2070BA)    C.01.04

Keyword: Integration

One-line Description:

Integration unable to handle complex chromatogram for area sum integration event

Problem:
When creating area sum integration event, the 'off' time for one group and the 'on' time for the next group must be separated by a minimum of 0.06 minutes. In a complex chromatogram this is a significant loss of data.
Temporary Solution:
n/a
Fix Information:
Fixed in revision C.01.04

KPR#: 1642    Product: Chemstation     ChemStation (G2070BA)    C.01.04

Keyword: Integration

One-line Description:

Data lost when using area sum integration event, must have 0.06 minute space between groups

Problem:
When creating area sum integration event, the 'off' time for one group and the 'on' time for the next group must be separated by a minimum of 0.06 minutes. In a complex chromatogram this is a significant loss of data.
Temporary Solution:
n/a
Fix Information:
 Fixed in revision C.01.04

KPR#: 1694    Product: Chemstation     ChemStation (G2070BA)    C.01.04

Keyword: Method

One-line Description:

Method marked Dirty on initial "Specify Report ..." with DEF_GC.M

Problem:
This problem is reproducible if you load DEF_GC.M. Once loaded, execute 'Edit Entire Method' and check 'Data Anylsis'. Close all dialogs with OK without doing changes. Then method becomes dirty [Error!].

Once the modifed DEF_GC.M is saved, the problem does not occur.

Also happens in DA View. If DEF-GC.M is loaded and "Specify report" dialog is entered and closed without a change, then the Method dirty flag is set.
Temporary Solution:
n/a
Fix Information:
n/a

KPR#: 1011    Product: Chemstation     ChemStation (G2070BA)    C.01.04

Keyword: Reports

One-line Description:

ChemStation does not report resolution as defined in European Pharmacopoeia

Problem:
ChemStation does not report resolution as defined in European Pharmacopoeia. EP requires the following formula to be used:
R = 1.18 * (Tr(b)-Tr(a)) / W50(b)+W50(a)


Only the following definitions are available in ChemStation for resolution:

USP, ASTM (in Extended Performance Report):
R = 2 * (Tr(b)-Tr(a)) / WB(b)+WB(a)

Half-width method as used in Performance Report:
R = (2.35/2) * (Tr(b)-Tr(a)) / W50(b)+W50(a)

JP (in Report Layout view only):
R = 1.18 * (Tr(b)-Tr(a)) / W50(b)+W50(a)
Temporary Solution:
In Report Layout view, use Resolution (JP) which uses the same formula.
Fix Information:
Fixed in revision C.01.04

KPR#: 1440    Product: Chemstation     ChemStation (G2070BA)    C.01.04

Keyword: Reports

One-line Description:

acquisition parameters missing in the Method section of the sequence summary report when Unique Folder Creation is switched off

Problem:
The acquisition parameters missing in the Method section of the sequence summary report when Unique Folder Creation is switched off. Only the run-time checklist, method information and data analysis parameters are reported.
Temporary Solution:
n/a
Fix Information:
Fixed in revision C.01.04

KPR#: 1679    Product: Chemstation     ChemStation (G2070BA)    C.01.04

Keyword: Reports

One-line Description:

sample name not reported with Intelligent Reporting in non-result set data that have been acquired prior to C.01.01

Problem:
The sample name is not reported when using Intelligent Reporting when working on data that have been required in revisions prior to OpenLAB CDS ChemStation Edition C.01.01.
Temporary Solution:
Use the "result set migration" tool of the Chemstation, to migrate the sequence to container format.
Fix Information:
 Fixed in revision C.01.04

KPR#: 1703    Product: Chemstation     ChemStation (G2070BA)    C.01.04

Keyword: Reports

One-line Description:

ChemMain.exe can crash when a long file path is used

Problem:
The file path in ChemStation has a hard coded limit of 256 characters.
In an ECM environment there are also hidden characters added to ensure unique file naming. As a consequence the complete lenth for  +  +  +  must not exceed a length of 175 characters.
Temporary Solution:
Make sure that the path length of a report from  -->  C:\Chem32\... to  ...pdf  <-- does not exceed 175 characters.
Fix Information:
No fix available.

KPR#: 1317    Product: Chemstation     ChemStation (G2070BA)    C.01.04

Keyword: SEQ

One-line Description:

Mismatch for sequence completed time stamp in sequence and instrument logbook

Problem:
The time stamps for sequence completion may differ between the sequence logbook and the instrument logbook:
The sequence logbook lists the time after completion of the acquisition, while the instrument logbook displays the time after completion of the sequence summary report.
Temporary Solution:
n/a
Fix Information:
Fixed in revision C.01.04

KPR#: 1636    Product: Chemstation     ChemStation (G2070BA)    C.01.04

Keyword: SEQ

One-line Description:

Headspace or PAL Sampler sequence with methods saved to a location other than C:\ does not run

Problem:
Save method on D drive, and start sequence with headspace or PAL Sampler.
    GC stays in "Prep Run".
    headspace and PAL Samplers do not work.

Headspace or PAL Sampler does not start when method is saved to a location other than C:\

This defect affects all of the B series headspace and PAL Sampler drivers.
Temporary Solution:
Workaround is to save and use methods from the C:\ drive.
Fix Information:
 Fixed in revision C.01.04

KPR#: 1709    Product: Chemstation     ChemStation (G2070BA)    C.01.04

Keyword: SEQ

One-line Description:

C.01.03: Partial Sequence Acquisition Does Not Work with Headspace with the Back Sequence Table

Problem:
It is not possible to run partial sequence acquisition using the back sequence table with Headspace.  

Exception is Found:
                                Exception
                                - Message:    Object reference not set to an instance of an object.
                                - Source:     AgtChemRCNetAdapter
                                - TargetSite: Void .ctor(PartialSequenceLine, SequenceLine, Int32, Int32, System.String, System.String)
                                - HelpLink:   
                                - Data:       System.Collections.ListDictionaryInternal
                                - StackTrace:
                                   at Agilent.ChemStation.RapidControl.DataHelper.RCSequenceEntryHelper..ctor(PartialSequenceLine partialSeqLine, SequenceLine seqLine, Int32 nMethodID, Int32 nPretreatId, String OnSampleIDMismatch, String strInjectorID)
                                   at Agilent.ChemStation.RapidControl.DataHelper.RCSequenceHelper.AddEntry(PartialSequenceLine partialSeqLine, SequenceLine seqLine, Int32 nMethodID, Int32 nPretreatID, String strInjectorID)
                                   at Agilent.ChemStation.RapidControl.AgtChemRCNetAdapter.FullSequence.CreateCollections(PartialSequenceTable partialSeqTable, SequenceContext sequenceContext, MethodCollection MethodCollection, PretreatCollection PretreatCollection, RCSequenceHelper SequenceHelper, Boolean bPreventMethodResolution)
                                   at Agilent.ChemStation.RapidControl.AgtChemRCNetAdapter.FullSequenceDownload.StartFullSequenceMode(String sampleIDMismatch)
                                   at Agilent.ChemStation.RapidControl.AgtChemRCNetAdapter.ControlAdapter.DoPreSequencePhase()


Investigation:  

The exception is the reason why the sequence is not downloaded to the instrument.  It is related to back sequences. Due to a minor mistake, the attempt to resolve a partial Back-sequence line fails, because the ChemStation tries to resolve it from the Front-Sequence Table.
Temporary Solution:
Workaround is to use Front-Sequences only.
Fix Information:
 Fixed in revision C.01.04

KPR#: 1713    Product: Chemstation     ChemStation (G2070BA)    C.01.04

Keyword: SEQ

One-line Description:

Opening the sequence table, click on a cell, make no changes, and select OK causes sequence to stop with 7697A Headspace Sampler

Problem:
If  the sequence table is opened during a running sequence and the user clicks on a cell and makes no changes, there is a lock placed on the vial so that the instrument doesn't put a vial in the oven. If the user makes no changes and hits OK then the Chemstation never continues past the locked vial.
Temporary Solution:
If the user makes a change to the sequence or hits the cancel button then the sequence works correctly.
Fix Information:
 Fixed in revision C.01.04

KPR#: 1736    Product: Chemstation     ChemStation (G2070BA)    C.01.04

Keyword: SEQ

One-line Description:

Partial sequence with methods saved to a location other than C:\ does not run with headspace and PAL Samplers

Problem:
A partial sequence run with methods saved to a location other than C:\ does not run. The logbook says it is running but the headspace or PAL Samplers do not run.
Temporary Solution:
n/a
Fix Information:
 Fixed in revision C.01.04

KPR#: 1695    Product: Chemstation     ChemStation (G2070BA)    C.01.04

Keyword: User Interface

One-line Description:

Change Access Level... is not available from menu bar

Problem:
Change Access Level... is grayed out and not available from the menu bar.
Temporary Solution:
n/a
Fix Information:
n/a

KPR#: 1740    Product: Chemstation     ChemStation (G2070BA)    C.01.04

Keyword: User Interface

One-line Description:

The logged-on user is not displayed in lock dialog

Problem:
Set authentication to Internal and create an 'everything' user
Log into Control Panel and open a 6890 session, which fails to initialize
This eventually times out and the lock dialog appears.  The logged-on user is not displayed in lock dialog. Do not know who is logged into Control Panel.
Temporary Solution:
The root cause is that the full name is not set.
Fix Information:
Fixed in revision C.01.04.